Configuration
Leakage test
The following tests are initiated during the leakage
test:
–
Compliance Test
–
System Leak Test
–
Ventilator Leak Test
–
Safety Relief Valves Test
1
Press the
Leak /ComplTest
softkey (A).
2
Follow the instructions on the screen.
After completion of the tests, the results (B) are
displayed on the screen.
3
To return to the start screen, press the rotary
knob.
Results of the compliance test
This test determines the system compliance
including breathing system, breathing hoses, filter,
and Y-piece.
The system compliance is required to ensure in
Volume Control
mode that the applied tidal
volume corresponds to the set tidal volume.
The value of the compliance is displayed on the
Standby
screen.
Results of the ventilator leakage test
The test of the ventilator leakage can have the
following results:
25463
21984
Standby
A
Standby
B
System compliance
[mL/cmH
2
O]
Displayed result
[mL/cmH
2
O]
≤
6.5
Measured value and
PASSED
Ventilator leakage
[mL/min]
Displayed result
[mL/min]
150
Measured value and
PASSED
151 to 250
Measured value and
FAILED
>250
>250 and
FAILED