Alarms
120
Instructions for use Fabius plus SW 3.n
Example for sounding the alarm tone when
several alarms are present
Suppressing the alarm tone
The alarm tone can be suppressed for a maximum
of 2 minutes.
Press the
(A) key.
The LED in the key (A) lights up.
In the status bar (B), the symbol
and the
remaining time for the alarm tone suppression are
displayed.
During the alarm tone suppression, only the new
alarms are acoustically signaled whose alarm
priority or internal priority number is higher than
the suppressed alarm, see chapter ''Alarm –
Cause – Remedy'' on page 160.
Reactivating the suppressed alarm tone
Press the
(A) key again.
The LED in the key (A) goes out.
Adjusting the alarm limits
If an alarm is triggered due to a value falling below
or exceeding the alarm limit, it might be necessary
to adjust the standard alarm limits. More
information for setting the alarm limits can be
found in chapter ''Changing the alarm limits'' on
page 141.
Setting the alarm limits in current ventilation mode:
Press the
key.
More information can be found in chapter
''Monitoring'' on page 121.
Priority of the existing alarm
Priority of the new alarm
Reaction from Fabius
(!!!) WARNING
(!!!) WARNING
– Alarm tone sequence starts
from the beginning.
(!!!) WARNING
(!!) CAUTION
– Alarm tone sequence for the
existing alarm is not inter-
rupted.
– No acoustic alarm signal for
the new alarm
(!!) CAUTION
(!!) CAUTION
– Alarm tone sequence starts
from the beginning.
(!!) CAUTION
(!!!) WARNING
– Alarm tone sequence for the
new alarm is started.