
Instructions for use Fabius plus SW 3.n
199
Technical data
Accuracy
Pressure limitation (
PMAX
)
±5 cmH
2
O (±5 hPa) of the setting
Tidal volume (
VT
)
±5 % of the setting or 20 mL, depending on which
value is higher (discharged to atmosphere, no
compliance correction)
Respiratory rate (
Freq
)
±1 bpm (±1 1/min) of the setting or ±5 %, depend-
ing on which value is higher
Ratio of inspiratory time to expiratory time (
TI:TE
)
±5 % of the setting
Inspiratory pause (
TIP:TI
)
±25 % of the setting
Positive end-expiratory pressure (
PEEP
)
±2 cmH
2
O (±2 hPa) or ±20 % of the setting,
depending on which value is higher
Inspiratory pressure (
PINSP
)
±2 cmH
2
O (±2 hPa) or ±20 % of the setting,
depending on which value is higher
Overpressure safety valve
75 ±5 cmH
2
O (75 ±5 hPa)
Underpressure safety valve (inlet valve for ambi-
ent air)
-7.5 to -9 cmH
2
O (-7.5 to -9 hPa)
Minimum pressure limit
-9 cmH
2
O (-9 hPa)
Measuring the system compliance
0.2 to 6.0 mL/cmH
2
O (0.2 to 6.0 mL/hPa)
±0.2 mL/cmH
2
O (±0.2 mL/hPa) or ±10 % of the
actual compliance depending which value is
higher