Instructions for use Fabius plus SW 3.n
Switching the volume alarms on and off
1
Press the
Volume
Alarms
ON
/
OFF
softkey
(A).
The key label changes from
Volume Alarms
ON
to
Volume Alarms
OFF
.
Instead of the upper and lower alarm limits, the
symbol appears, indicating the alarm is
deactivated.
The volume alarms are deactivated.
Automatic setting of the pressure
threshold
1
Press the
Auto
Set
softkey (A).
The pressure threshold for the peak pressure
(
PEAK
) is set to 4 cmH
2
O (hPa) below the
current plateau pressure (
PLAT
).
25505
NOTE
The function
Volume Alarms
ON
/
OFF
is
available in the standard view of the
ManSpont
mode. If the
key is pressed in
ManSpont
mode, the softkey
Volume
Alarms
ON
/
OFF
is
not displayed.
Volume Control
A
25505
NOTE
The pressure threshold must not be below
5 cmH
2
O (5 hPa) or above 30 cmH
2
O (30 hPa).
NOTE
If no current measured value is available for the
plateau pressure (
PLAT
), pressing the softkey
has no effect.
NOTE
In
SIMV/PS
mode, the pressure threshold
depends on the pressure of the mandatory
breaths.
A
Volume Control