Instructions for use Fabius plus SW 3.n
79
Assembly and preparation
Connecting the sensors and measurement lines
Connecting the O
2
sensor
1
Insert the O
2
sensor (A) into the opening of the
inspiratory valve on the compact breathing
system.
2
Connect the plug (B) of the O
2
sensor cable
with the connection marked with
on the
rear of the device.
Connecting the pressure sensor
1
Connect the pressure measurement hose (C)
to the corresponding connection on the bottom
of the compact breathing system.
Make sure that the pressure measurement
hose is not pinched.
2
Connect the other end of the pressure
measurement hose with the bacterial filter (B).
3
Plug in the bacterial filter into the connection
(C) marked with
on the rear of the device.
25444
A
B
25446
A
B
C