E.
Place the test weight on the platform. Adjust the displayed weight to match the
test weight using the UP or DOWN arrows.
F.
When the displayed weight is correct, press the UNITS button to save the calibration
- the display will show "Ent SP" to signify that the span has been successfully
calibrated.
G.
Open SW1 (SW2 #9 for M model) to exit the setup mode and return to normal
operation. Reassemble the indicator.
See Section 6 for more information on the setup mode, see Section 7 for a more detailed
description of the digital calibration procedure.
If an error message was encountered at any time during the calibration, the calibration was
not successful. See section 11 for a list of error codes and their meaning.
Section 2 Quick Setup Guide
7