Not Ready Symptoms
4
Troubleshooting and Maintenance Guide
79
A Gas Exceeds Pressure Setpoint or Flow
If a gas exceeds its pressure or flow setpoint, do the following:
•
Decrease the split ratio.
•
Replace the split vent filter.
•
Verify that the correct liner is selected.
•
Check the gold seal for contamination.