200
Troubleshooting and Maintenance Guide
13
Analyzer
Overview
The analyzer (
) is the heart of the MSD. It ionizes the
sample, filters the ions, and detects them. The sample
components exiting the GC column flow into the ion source. In
the ion source, the sample molecules are ionized and
fragmented. The resulting ions are repelled from the ion source
into the quadrupole mass filter. The mass filter allows selected
ions to pass through the filter and strike the detector. The
detector generates a signal current proportional to the number
of ions striking it.
The analyzer is attached to the vacuum side of the side plate.
The side plate is hinged for easy access. The ion source and the
mass filter are independently heated. Each is mounted inside a
radiator for correct heat distribution.
Each of the parts of the analyzer is discussed in the following
material.
The analyzer has four basic components
The analyzer consists of the following components (
):
•
Ion source
•
Mass filter
•
Detector
•
Heaters and radiators
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...