To Replace the Primary Fuses
Materials needed
•
Fuse, T8 A, 250 V (2110-0969) – 2 required
•
Screwdriver, flat-blade (8730-0002)
The most likely cause of failure of the primary fuses is a
problem with the foreline pump. If the primary fuses in your
MSD fail, check the foreline pump.
Procedure
1
Vent the MSD and unplug the power cord from the electrical
outlet.
If one of the primary fuses has failed, the instrument will
already be off, but for safety you should switch off the
instrument and unplug the power cord. It is not necessary to
allow air into the analyzer chamber.
2
Turn one of the fuse holders (
) counterclockwise
until it pops out. The fuse holders are spring loaded.
3
Remove the old fuse from the fuse holder.
4
Install a new fuse in the fuse holder.
5
Reinstall the fuse holder.
WA R N I N G
Never replace the primary fuses while the instrument is
connected to a power source.
WA R N I N G
If you are using hydrogen as a GC carrier gas, a power failure may
allow it to accumulate in the analyzer chamber. In that case,
further precautions are required. See
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...