![Agilent Technologies 5975T LTM GC/MSD Скачать руководство пользователя страница 121](http://html1.mh-extra.com/html/agilent-technologies/5975t-ltm-gc-msd/5975t-ltm-gc-msd_troubleshooting-and-maintenance-manual_2867607121.webp)
General Maintenance
10
Troubleshooting and Maintenance Guide
121
To Drain the Foreline Pump
Standard foreline pump only
Materials needed
•
Book or other solid object approximately 5 cm thick
•
Container for catching old pump oil, 500 mL
•
Gloves, oil- and solvent-resistant
•
Screwdriver, flat-blade, large (8730-0002)
•
Hex key, 5-mm (8710-1838)
Procedure
1
Vent the instrument.
2
If necessary, slide the foreline pump to a safe, accessible
location.
The foreline pump may be located on the floor or on the lab
bench next to the GC/MSD.
3
Place a book or other object under the pump motor to tilt it
up slightly. Remove the fill cap. See
.
4
Place a container under the drain plug.
5
Remove the drain plug. Allow the pump oil to drain out. The
oil drains faster if it is still warm.
6
Replace the drain plug after draining the oil.
7
Refill the foreline pump until the oil level is between the two
fill marks in the site window.
8
Replace the fill cap.
WA R N I N G
The foreline pump can cause burns if touched when operating. It
has a safety shield to prevent the user from touching it.
WA R N I N G
The foreline pump may be hot.
WA R N I N G
The old pump oil may contain toxic chemicals. Treat it as
hazardous waste.
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...