22
Troubleshooting and Maintenance Guide
1
Introduction
•
Eliminate from your laboratory as many ignition sources as
possible (open flames, devices that can spark, sources of
static electricity, etc.).
•
Do not allow hydrogen from a high pressure cylinder to vent
directly to atmosphere (danger of self-ignition).
•
Use a hydrogen generator instead of bottled hydrogen.
Operating precautions
•
Turn off the hydrogen at its source every time you shut
down the instrument.
•
Turn off the hydrogen at its source every time you vent the
MSD (do not heat the capillary column without carrier gas
flow).
•
Turn off the hydrogen at its source every time shutoff valves
in an MSD are closed (do not heat the capillary column
without carrier gas flow).
•
Turn off the hydrogen at its source if a power failure occurs.
•
If a power failure occurs while the GC/MSD system is
unattended, even if the system has restarted by itself:
1
Immediately turn off the hydrogen at its source.
2
Turn off the instrument and allow it to cool for 1 hour.
3
Eliminate
all
potential sources of ignition in the room.
4
Open the vacuum manifold to atmosphere.
5
Wait at least 10 minutes to allow any hydrogen to
dissipate.
6
Start up the instrument as normal.
When using hydrogen gas
check the system for leaks to prevent
possible fire and explosion hazards based on local
Environmental Health and Safety (EHS) requirements. Always
check for leaks after changing a tank or servicing the gas lines.
Always make sure the vent line is vented into a fume hood.
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...