112
Troubleshooting and Maintenance Guide
10
General Maintenance
•
GC/MSD interface
•
Analyzer parts
•
Vacuum pumps
•
Inlet
•
LTM column
•
Guard column heated enclosure
The GC inlets, guard column heated enclosures, and LTM
column also operate at very high temperatures. Use the same
caution around these parts.
Chemical residue
Only a small portion of your sample is ionized by the ion source.
The majority of any sample passes through the ion source
without being ionized. It is pumped away by the vacuum system.
As a result, the exhaust from the foreline pump will contain
traces of the carrier gas and your samples. Exhaust from the
standard foreline pump also contains tiny droplets of foreline
pump oil.
An oil trap is supplied with the standard foreline pump. This
trap stops
only
pump oil droplets. It
does not
trap any other
chemicals. If you are using toxic solvents or analyzing toxic
chemicals, do not use this oil trap. For all foreline pumps, install
a hose to take the exhaust from the foreline pump outdoors or
into a fume hood vented to the outdoors. For the standard
foreline pump, this requires removing the oil trap. Be sure to
comply with your local air quality regulations.
WA R N I N G
Never touch these parts while your instrument is on. After the
instrument is turned off, give these parts enough time to cool
before handling them.
WA R N I N G
The foreline pump can cause burns if touched when operating. It
has a safety shield to prevent the user from touching it.
WA R N I N G
The oil trap supplied with the standard foreline pump stops only
foreline pump oil. It does not trap or filter out toxic chemicals. If
you are using toxic solvents or analyzing toxic chemicals, remove
the oil trap and install a hose to take the foreline pump exhaust
outside or to a fume hood.
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...