![Agilent Technologies 5975T LTM GC/MSD Скачать руководство пользователя страница 71](http://html1.mh-extra.com/html/agilent-technologies/5975t-ltm-gc-msd/5975t-ltm-gc-msd_troubleshooting-and-maintenance-manual_2867607071.webp)
Chromatographic Symptoms
3
Troubleshooting and Maintenance Guide
71
91, 92
Toluene or xylene
Cleaning solvent
105, 106
Xylene
Cleaning solvent
151, 153
Trichloroethane
Cleaning solvent
69
Foreline pump oil or PFTBA
Foreline pump oil vapor or calibration
valve leak
73, 147, 207, 221, 281, 295, 355, 429
Dimethylpolysiloxane
Septum bleed or methyl silicone
column bleed
149
Plasticizer (phthalates)
Vacuum seals (O-rings)
damaged by high temperatures, vinyl
gloves
Peaks spaced 14
m/z
apart
Hydrocarbons
Fingerprints, foreline pump oil
Table 4
Common contaminants (continued)
Ions (
m/z
)
Compound
Possible source
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...