General Maintenance
10
Troubleshooting and Maintenance Guide
127
To Purge the Calibration Valves
EI calibration valve
After adding new PFTBA to the calibrant vial, you must purge
the air out of the vial and valve.
1
If the vacuum gauge controller is on, turn it off.
2
In
Tune and Vacuum Control
view, select
Purge Calibrant Valve
under the
Vacuum
menu.
This will open the CI calibration valve for several minutes
with all analyzer voltages turned off.
C A U T I O N
After removing a calibrant vial, you
must
purge the calibration valve.
Failure to do so will result in damage to the filaments and the
electron multiplier.
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...