132
Troubleshooting and Maintenance Guide
10
General Maintenance
To Remove the Micro-Ion Vacuum Gauge
Procedure
1
Vent the instrument.
2
Remove the MSD back cover.
3
Disconnect the cable on the back of the Micro-Ion vacuum
gauge.
4
Unscrew the large thumbnut on the gauge clamp.
5
Remove the long screw from the clamp.
6
While supporting the gauge body, remove the clamp from the
mounting flange.
7
Remove the gauge.
8
If you will not be replacing the gauge soon, install the
blanking plate provided with the gauge and secure it with
the clamp, screw, and thumbnut.
9
Replace the rear MSD cover. See
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...