Chromatographic Symptoms
3
Troubleshooting and Maintenance Guide
47
Sample Decomposition in Inlet/Missing Peaks
•
Lower the inlet temperature.
•
Check for air or water in the carrier gas; verify gas purity
and functionality of traps.
•
Verify that the liner is appropriate for the sample being run.
•
Perform complete inlet maintenance: Replace all
consumable parts and bake out the inlet.
•
Install a deactivated liner.
•
Check for leaks at the septum, liner, and column fittings.
(See
•
Install an Agilent Direct Connect liner.
•
Use a pulsed pressure method for quicker sample transfer to
column.
•
Bake out the inlet.
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...