To Maintain the Dry Vacuum Pump
Materials needed
•
Replacement tip seal kit (IDP3TS)
•
Wrench, open-end, 1/4-inch × 5/16-inch (8710-0510)
•
Razor blade or side-cutting pliers
•
Compressed air (optional)
Procedure
1
Vent the instrument.
2
If necessary, slide the foreline pump to a safe, accessible
location.
The foreline pump may be located on the floor or on the lab
bench next to the GC/MSD.
3
Disconnect the pump from electrical power.
4
Remove the four M5 socket head bolts from the front
cowling, disconnect the electrical connector, and set the
cowling aside. See
.
5
Remove the four M5 bolts from the outboard housing and
take it axially off the frame.
6
Remove the worn tip seals and discard.
7
If compressed air is available, blow any remaining tip seal
debris from the scroll parts. If seal debris is attached to the
sides, use a razor blade to scrape it off.
8
Remove the new tip seals from the packaging. There should
be two tip seals in the kit: one for the orbiting scroll and one
for the outboard housing scroll.
9
Install the correct tip seal into the groove of the orbiting
scroll, and trim about 3 mm off the seal from the outer end
of the groove. See
10
Install the correct tip seal into the groove of the outboard
housing scroll and trim as in
WA R N I N G
The foreline pump can cause burns if touched when operating. It
has a safety shield to prevent the user from touching it.
WA R N I N G
The foreline pump may be hot.
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...