Chromatographic Symptoms
3
Troubleshooting and Maintenance Guide
53
Poor sensitivity
•
Incorrect tuning
•
Tune file that does not match the type of analysis
•
Repeller voltage is too low
•
Incorrect temperatures (column, GC/MSD interface, ion
source, or mass filter)
•
Incorrect sample concentration
•
Leaking GC inlet*
•
Dirty GC inlet
•
Incorrect split ratio
•
Purge off time in splitless mode is too short
•
Excessive pressure in the MSD
•
Dirty ion source
•
Air leak
•
Poor filament operation
•
Detector (HED electron multiplier) is not working correctly
•
Incorrect mass filter polarity
* This could cause a fault condition in the GC that would prevent
the GC from operating.
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...