![Agilent Technologies 5975T LTM GC/MSD Скачать руководство пользователя страница 187](http://html1.mh-extra.com/html/agilent-technologies/5975t-ltm-gc-msd/5975t-ltm-gc-msd_troubleshooting-and-maintenance-manual_2867607187.webp)
Vacuum System
12
Troubleshooting and Maintenance Guide
187
The standard foreline pump is a two-stage rotary-vane pump.
An optional dry pump is also available. The pump turns on
when the instrument power is turned on. The foreline pump has
a built-in antisuckback valve to help prevent backstreaming in
the event of a power failure.
The foreline pump can be placed on the bench beside the
instrument (with the exhaust outlet to the rear) or on the floor
below the instrument.
An oil trap (not shown) is available for the standard wet pump
that can be used to filter pump oil out of the foreline pump
exhaust. This trap stops
only
pump oil. Do not use the trap if
you are analyzing toxic chemicals or using toxic solvents.
Instead, install an 11-mm id hose to remove the exhaust from
your lab.
A window (sight glass) in the front of the standard foreline
pump shows the level of the foreline pump oil. There are two
marks next to the window. The level of the pump oil should
never be above the upper mark or below the lower mark. If the
level of pump oil is near the lower mark, add foreline pump oil.
WA R N I N G
The oil trap supplied with the standard foreline pump stops only
foreline pump oil. It does not trap or filter out toxic chemicals. If
you are using toxic solvents or analyzing toxic chemicals, remove
the oil trap and install a hose to take the foreline pump exhaust
outside or to a fume hood.
C A U T I O N
Do not place the foreline pump near any equipment that is sensitive
to vibration.
C A U T I O N
The ballast control knob controls the amount of air allowed into the
pump. Keep the ballast control closed (fully clockwise) at all times,
except when ballasting the pump.
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...