General Maintenance
Maintaining the Vacuum System
Periodic maintenance
As listed earlier in
, some maintenance tasks for the
vacuum system must be performed periodically. These include:
•
Checking the foreline pump fluid (every week)
•
Checking the calibration vial (every 6 months)
•
Ballasting the foreline pump
•
Replacing the foreline pump oil every 6 months
•
Tightening the foreline pump oil box screws (first oil change
after installation, standard foreline pump)
•
Checking the dry foreline pump diaphragms (typically every
3 years)
Failure to perform these tasks as scheduled can result in
decreased instrument performance. It can also result in damage
to your instrument.
Other procedures
Tasks such as replacing a Micro-Ion vacuum gauge should be
performed only when needed. See
or the online help
in the GC/MSD ChemStation software for symptoms that
indicate this type of maintenance is required.
More information is available
If you need more information about the locations or functions of
vacuum system components, See
Most of the procedures in this chapter are illustrated with video
clips on this 5975 Series MSD User Information DVD.
Содержание 5975T LTM GC/MSD
Страница 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Страница 10: ...10 Troubleshooting and Maintenance Guide...
Страница 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Страница 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Страница 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Страница 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Страница 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Страница 150: ...150 Troubleshooting and Maintenance Guide 10 General Maintenance Figure 16 Replacing filaments Filament 1 Filament 2...
Страница 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Страница 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...
Страница 239: ...Parts 15 Troubleshooting and Maintenance Guide 239 Figure 39 LCP replacement parts LCP assembly and fascia...