Carl Zeiss
Configuration of the Light Path
Lightsheet Z.1
32
000000-1790-528
02/2013
•
The light path has been changed automatically, removing the laser blocking filter from the beam path,
so stray light from the activated laser can be seen as a live image on the left hand side. Hence any
bright or blocking structures, like the sample holder or the sample, are visible. If this is the case, the
sample must be moved (optimally in the y direction), resulting in an evenly illuminated field of view.
•
Click
Next
.
•
You can follow the status of the adjustment on the progress bar.
Summary of Contents for Lightsheet Z.1
Page 1: ...Lightsheet Z 1 Operating Manual February 2013 ZEN 2012 black edition ...
Page 4: ......
Page 170: ......
Page 427: ...Lightsheet Z 1 Overview ...