Lightsheet Z.1
Carl Zeiss
02/2013
000000-1790-528
37
3.7
Installation of the Incubation Modules
The incubation of the Lightsheet Z.1 is configuration dependent. It is used to control and monitor
temperature and CO
2
supply to the sample environment. The
TempModule S1
is required for all of the
various configurations.
−
The configuration dependent modules should be placed on a level surface next to the system and
are stacked on top of each other (Fig. 31). For further information see the individual component
manuals from PeCon.
The installation of incubation modules on the
Lightsheet Z.1 can be divided into two parts. One
deals with the actual assembly of hardware
components of the incubation modules and their
connections to the system. The second part deals
with the registration of the incubation modules
using specific software tools.
3.7.1
Heating Components
There are two different heating components available:
(1)
Heating Block
Sample Chamber
Lightsheet Z.1 enables to heat the sample chamber from ambient temperature to 42 °C.
(2)
Peltierblock
Sample Chamber Lightsheet Z.1 permits heating and cooling of the sample chamber in the range from
10 °C to 42 °C.
(1) Heatingblock
Sample Chamber Lightsheet Z.1 is controlled by the TempModule S1 providing
electrical power supply and closed-loop temperature control. The system allows to heat the sample
chamber from ambient temperature to 42 °C
.
Connect the Heating Block LSFM and the TempModule S via connecting cable. For proper electrical
connection please follow the connection diagram in the individual component manuals from
PeCon.
Fig. 31
Incubation modules stacked on top
of each other