Lightsheet Z.1
Carl Zeiss
02/2013
000000-1790-528
31
3.4
Removing and Inserting the Sample Chamber
The sample chamber can be removed from the system. For removal, proceed as follows:
−
Open the front system cavity door (Fig. 25/
1
) of the main system module Lightsheet Z.1.
−
Remove the liquid from the sample chamber (Fig. 25/
6
) using the hose with the corresponding
syringe (Fig. 25/
9
).
−
Ensure that no sample protrudes into the sample chamber. Remove the sample, if necessary,
through the upper sample opening (Fig. 25/
2
) below the upper system cavity door (Fig. 25/
3
).
−
If necessary, remove the connections for the incubation device (Fig. 25/
8
).
−
Loosen the screw (Fig. 25/
4
) on the sample chamber and pull the sample chamber out by the grip
5
).
−
Inserting the sample chamber:
−
Take hold of the sample chamber by the lower grip (Fig. 25/
5
) and push the sample chamber
6
) into the guide rails (Fig. 25/
7
).
−
Tighten the screw (Fig. 25/
4
) without exerting force. The sample chamber has been correctly
inserted when its lower edge is flush with the guide rail.
−
Plug in the cable to the incubation unit (Fig. 25/
8
) as necessary.
1
Cover for capillaries (3 mm opening)
2
Cover for syringes (7 mm opening)
3
Sample chamber body with dove tail slide
Fig. 24
Placing cover on top of the sample chamber body