![background image](http://html1.mh-extra.com/html/zeiss/lattice-lightsheet-7/lattice-lightsheet-7_instructions-manual_3406249035.webp)
SAFETY INFORMATION
Lattice Lightsheet 7
User Interface - Mounting and Dismounting Components
ZEISS
04/2021 V_01
000000-2324-960
35
The rear panel contains all plugs for electrical connections (Fig. 19).
The following electrical connections must be made:
•
Connect the power plug (Fig. 19/
14
) via the power cable to the mains power 3-point connector.
•
Connect channel 1 / heated lid, channel 2 / heated tray, and channel 3 / heated base (Fig. 19/
9, 10, 11
)
with the gas supply plug line plug (Fig. 2/
12
) of the Lattice Lightsheet 7.
There is no need to connect the spare (heated lens) plug (Fig. 19/
12
).
•
Connect the USB (Fig. 19/
8
) to an USB port of the system computer.
•
Connect the CAN bus 2 (Fig. 19/
7
) to the Gas CAN Bus 1 plug ((Fig. 2/
11
) of the Gas Mixer. On the CAN
Bus 1 socket insert a CAN Bus blind plug..
1 Power
control LED for power channel
6 CAN Bus 2
Can Bus connection to Gas Mixer
2
Lid
control LED for Lid channel
7
CAN Bus 1
Can Bus blind plug
3
Tray
control LED for tray (insert) channel
8
USB
for connecting to computer
4
Base
control LED for base channel
9
Heated lid
channel 1
5
Spare
control LED for free channel
10
Heated tray
channel 2
11 Heated base
channel 3
12 Spare
channel 4 (not needed)
13 On/Off
on/off switch
14 Power
external power supply
15 Temp sens
for connection to external sensor (optional)
Fig. 19