SAFETY INFORMATION
ZEISS
Laser Lifetimes
Lattice Lightsheet 7
20
000000-2324-960
04/2021 V_01
1.10
Laser Lifetimes
The following table shows the expected laser life times:
Laser
Type
Expected Life
time (hrs)
Class
Nominal
Power
Power at the end of
expected life time (hrs)
1 Laser 488 nm
Diode
10000
3B
10 mW
8 mW
2 Laser 561 nm
Diode (SHG)
5000
3B
10 mW
8 mW
3 Laser 640 nm
Diode
10000
3B
5 mW
4 mW
1.11
Notes on Handling the Computer, Data Media and Data Transfer
The computer used as standard in your Lattice Lightsheet 7 system is a high-end computer with WINDOWS
operating system.
As standard, your computer has one hard disk drive and one CD-ROM drive.
On delivery of your Lattice Lightsheet 7 system, please ensure that the operating system is
installed including initialization and start file setup and that the ZEN program is also installed
When working with the hard disk, it is important to know that the more data it contains, the
slower its operation will become. Therefore, data that you do not need permanently should be
stored on a DVD, a CD-ROM or an external HD drive.
When handling CDs, CD-ROMs or DVDs, do not touch the data side of the disc. Do not apply
paper labels or write on any part of the disc, data side or label side. Wipe dust or fingerprints on
the disc with a soft cloth from the center to the edge. Do not use benzene, thinner, record cleaner,
or antistatic clothes. This may damage the disc. Do not expose the disc to direct sunlight or high
temperatures. Backup your data on a regular basis. Do not install any other software without
consulting your ZEISS representative.
Never turn your computer off before exiting the ZEN program and shutting down the WINDOWS
operating system. Otherwise, program files and/or data files may get lost.
The Lattice Lightsheet 7 computer is a system computer which controls a high-end microscope
system. The computer must not to be used as a general workstation with frequent updates of
windows or security patches. The ZEN software is released with a specific windows version and
patch and this must not be changed unless approved by Carl Zeiss Microscopy.
Necessary data transmission rates of the network are dependent on the tasks to be performed. For standard
data transfer and online condition monitoring a data transmission rate of 1 MBit/s is sufficient. For Remote
Desktop Support a data transmission rate of at least 5 Mbit/s is required. To transfer big data sets we
recommend a data transmission rate of at least 10 Mbit/s.