SAFETY INFORMATION
Lattice Lightsheet 7
User Interface - Mounting and Dismounting Components
ZEISS
04/2021 V_01
000000-2324-960
29
1.13.2.2
Immersion Priming
The immersion supply system is depicted in Fig. 14. It is a water immersion system, so only water should be
used as immersion.
Do not use oil- or glycerol-based immersion media, as these more viscous solution will clog the tubing
system (Fig. 14/
4
).
1
Water immersion drop
6
3-way valve (arrow indicates flow direction)
2
Relay optics (meniscus lens)
7
Pump (arrow indicates flow direction)
3
Relay optics (meniscus lens holder)
8
Immersion container (water)
4
Hydraulic tubing
9
Luer-Lock stopcock
5
Luer-Lock coupling
10
Syringe
Fig. 14
Immersion supply system