OPERATING INSTRUCTIONS
ZEISS
Switching on and off the Lattice Lightsheet 7 System
Lattice Lightsheet 7
72
000000-2324-960
04/2021 V_01
3.1.2.5
Environmental Control (Optional)
Set the required temperature in the Temperature
expansion field of the Incubation tab (Fig. 51) of the
right tool area. You can select different temperatures
for
Lid
,
Tray
and
Base
.
We recommend starting with heating Lid and
Tray only and set the Lid temperature
2 degrees above the wanted temperature.
Set the required atmosphere in the
Atmosphere
expansion field of the Incubation tab (Fig. 51) of the
right tool area. You can select
O
2
%
,
CO
2
%
and
Hum%
(humidity).
We recommend starting with CO
2
levels 2%
higher than the wanted CO
2
percentage and a
humidity of 75%.
If heated to 37
o
C it is recommended to
replenish water immersion every 12 min. using
the auto immersion tool.
Fig. 51
Incubation tab of right tool area