JD700B Series User’s Guide
753
5
Optional
. To enable the external offset, complete the following steps:
Toggle the
External Offset
soft key and select
On
.
Enter a value between 0 and 60 by using the numeric keys.
Press the
dB
soft key.
Making a measurement
Procedure
1
Press the
MEASURE
hot key.
2
Depending your connected power sensor, do one of the following:
For JD736B, select the power range option:
Average Power
,
Peak Power
, or
Pulse Power
.
For JD734B, select
Peak Power
.
For JD732B, select
Average Power
.
For JD731B, select the power range option:
Forward Avg Power
,
Reverse Avg Power
,
Forward Peak Power
,
Pulse Power
, or
VSWR
.
For JD733A, select the power range option:
Forward Avg Power
,
Reverse Avg Power
,
Forward Peak Power
, or
VSWR
.
3
Press the
MEASURE SETUP
hot key.
4
Press the
Resolution
soft key, and then select the resolution option from the following choices:
0:
Displays the reading with no zero decimal place
1:
Displays the reading with one decimal place
2:
Displays the reading with two decimal places
5
To set a threshold and view the results with Pass/Fail indication, complete the following steps:
Press the
LIMIT
hot key.
Toggle the
Limit
soft key between
On
and
Off
to enable or disable the Pass/Fail indication.
Press the
High Limit
soft key, and then enter a threshold value.
Press the
dBm
unit soft key.
Press the
Low Limit
soft key, and then enter a threshold value.
Press the
dBm
unit soft key.
6
Optional
. Go to
SAVE/LOAD > Save
, and then select the save option from the choices available
for your measurement mode. See “Using save” on page 35 for more information.
Measurement example
You can measure and understand a trend of a system output in operation where output power constantly
varies depending on the amount of calls being processed by the BTS or Repeater. In addition, the Log graph
benefits you to easily examine the discontinuance of output power caused by the transmission problems or
defective connection.