Chapter 15
Using Cable and Antenna Analyzer
JD700B Series User’s Guide
527
Figure 317 2-port calibration connection with EZ-Cal in vector measurements
1)
2)
Procedure
1
Connect an extension cable, if necessary, to the
Reflection/RF Out
port of the instrument.
2
Connect your
EZ-Cal
to the other open end of the extension cable or to your instrument's
Reflection/RF Out
port directly.
3
Connect the USB cable to the side of the
EZ-Cal
and then to the
USB Host
port to power the
e-calibration kit as illustrated in Figure 317
-1
.
The instrument detects the EZ-Cal automatically and displays the EZ-Cal icon.
4
Press the
MEASURE SETUP
hot key.
5
Press the
Calibrate
soft key.
The on-screen instruction appears to guide you through the e-calibration.
6
When the O-S-L calibration is done, make sure to disconnect the EZ-Cal kit including its USB
connection.
7
Connect the
Reflection/RF Out
port and the
RF In
port for the Cable and Antenna Analyzer with
the test cable and the proper adapter as illustrated in Figure 317
-2
.
8
Press the
Continue
soft key to continue calibration.
The calibration progress bar appears.
After completion, the calibration status on the screen changes to
ON
.
Scalar measurement
illustrates the connection for calibration in Scalar measurements.
Figure 318 2-port calibration connection in Scalar measurements
Procedure
1
Press the
MEASURE SETUP
hot key.
2
Press the
Calibrate
soft key.
The on-screen instruction for the calibration appears.
3
Connect the
Cable and Antenna Analyzer Reflection/RF Out
port and the
Spectrum Analyzer
RF In
port with a test cable as illustrated in
4
Press the
Continue
soft key to continue calibration.
The calibration progress bar appears.