Chapter 10
Using CDMA Signal Analyzer
JD700B Series User’s Guide
371
Measurement example
Figure 220 CDP Table measurement with CDMA signal analyzer
Setting display
If the CDP table spans over to the next page(s), you can change pages up and down to view your
measurement results.
Procedure
1
Press the
TRACE/DISPLAY
hard key.
2
Press the
Code Order
soft key, and then select the code order option from the following choices:
Hadamard:
To assign the number based on Walsh code order and display the measured
code power according to the assigned number in sequence.
Bit Reverse:
Code number in binary bit is reversed relative to the Hadamard method.
3
Toggle the
Base Code Length
soft key and select the base Walsh code number:
64
or
128
.
4
Select
Page Up
or
Page Down
as needed to scroll the screen up and down.
Setting limit for modulation tests
Procedure
1
Press the
LIMIT
hot key.
2
Press the Modulation Test Limits soft key.
3
Select the test item(s) and set the limit(s) depending on your selected measurement mode:
To set the limit for
Select
Set
Frequency error
Frequency Error
High Limit, Low Limit
Time offset
Time Offset
High Limit, Low Limit
Waveform quality (Rho)
Rho
Low Limit
Carrier feed through
Carrier Feed Through
High Limit
Pilot power
Pilot Power
High Limit [Abs|Rel]