Chapter 15
Using Cable and Antenna Analyzer
JD700B Series User’s Guide
545
5
Press the
Autoset
soft key to let the instrument set the limit for each segment and display the line.
6
Optional
. To move the limit line, complete the following steps:
Select
Limit Up/Down
or
Limit Left/Right
.
Turn the rotary knob to move the line as desired. You can also manually enter a value.
7
Optional
. To edit the segment properties, complete the following steps:
Press the
Edit Limit
soft key.
Press the
Move
soft key and then turn the rotary knob to select the segment to edit.
Select the menu option, from the following choices:
To hide the line for the selected segment, toggle the
Line
soft key and select
Off
.
To add a new point, press the
Add Point
soft key.
To delete the selected point, press the
Delete Point
soft key.
To change the position, press the
Frequency
or
Amplitude
soft key, and then turn the
rotary knob to change the value as desired.
8
Optional
. Go to
SAVE/LOAD > Save
, and then select
Limit
to save the limit settings.
See “Using save” on page 35 for more information.