4-Point Bright Calibration
To capture 4-point bright calibration data:
1. Run the
VAKAP.exe
file.
Norm
C:\VCaptureSW\Acquisition\CEPH\WideCEPHG\Norm
Fast
C:\VCaptureSW\Acquisition\CEPH\WideCEPHG\Fast
2. On the dialog box, Select
Settings
and follow the steps to set exposure conditions:
1) Go to
[BrightCalibration]
and enter
4
for the NumbPoints.
2) Check if the tube voltage (kVp) and current (mA) are set as the table below. If
necessary, type each value according to the table.
Exposure Condition
Values
kVp0
780
mA0
970
kVp1
650
mA1
1000
3) Click
Save
.
①
②
③