4. Press and hold the X-ray exposure switch when you see the message “
HOLD THE
BUTTON
” (marked in the white box).
5. Release the switch when you see the message “
RELEASE THE BUTTON
” (marked in the
white box). If the system does not show this message, repeat
step 3
.
6-1. When you see a pop-up message notifying the end of auto CEPH auto-alignment, click OK
to finish.
6-2. When the system shows an error message, repeat
steps 1
to
5
until the CEPH auto-
alignment is completed.