Chapter 6.
CEPH Image Optimization
6.1.
CEPH Sensor Calibration
6.1.1.
Before CEPH Sensor Calibration
Before CEPH sensor calibration, install VAKCAP.com program.
Path
:
C:\VCaptureSW\Acquisition\Ceph\WideCeph\Norm\ folder
Ensure to save all existing CEPH sensor calibration folder before starting
a new calibration. You can find the previous folders at
C:\VCaptureSW\Acquisition\CEPH\WideCeph\Norm\CAL
Follow the steps below to prepare the system for a CEPH sensor calibration:
1. Remove the normal chin block.
2. Remove the CEPH ear rods and fold the nasal positioner to prevent the CEPH sensor from
blocking.
3. Disassemble the tube cases following the instruction in Chapter 1.
4. Rename the CAL folder for backup.
(
Path
:
C:\VCaptureSW\Acquisition\Ceph\WideCeph\Norm\CAL
)
If you are using an old collimator
,
ensure to reassemble the cu filter as illustrated below
.
Before reassembling Cu filter After reassembling Cu filter
New Collimator Information is as listed below:
Application date:
2021-08-20
Cut in S/N:
098-000155
Summary of Contents for A9 PHT-30CSS
Page 1: ...vatech A9 Technical manual Model PHT 30CSS Version 1 01 English...
Page 12: ...Page 11 of 244 1 2 1 Vertical Frame Part...
Page 22: ...Page 21 of 244 1 2 3 5 Generator DG 07E22T2...
Page 26: ...Page 25 of 244...
Page 27: ...Page 26 of 244...
Page 30: ...Page 29 of 244 1 3 3 Sensor Part Connection Diagram...
Page 31: ...Page 30 of 244 1 3 4 CEPH SENSOR Part Connection Diagram...
Page 32: ...Page 31 of 244 1 3 5 Connection Diagram Total assembly...
Page 43: ...Page 42 of 244 16 Loosen ten bolts as shown in the image below...
Page 112: ...Page 111 of 244 5 Click the V Vol button Click the Ready button...
Page 180: ...Page 179 of 244 10 Select 50 in the M box and click the center of the flat washer...
Page 209: ...Page 208 of 244 8 5 Appendix 8 5 1 Main MCU Board Parameters...
Page 244: ...vatech A9 www vatech com...