of
244
3. Click
Dark
to capture an image.
4. Check if the median value of the capture image satisfies the standards below.
Min
Max
Median value
700
2300
If the median value from the captured image does not meet the dark calibation standards,
replace the sensor and repeat the calibaration procedure.
The wrong median value from the captured image indicates that your
sensor is defective.