Page
192
of
244
Chapter 7.
CEPH Image Evaluation
7.1.
Lateral Mode Start Position Evaluation
7.1.1.
Lateral Mode Starting position standard
Parameter
LSL (Lower Specification Limit)
USL (Upper Specification Limit)
LAT start position
605
625
7.1.2.
Phantom Specifications
CEPH Phantom QUART Phantom
7.1.3.
Test Method
1. Run Console Software and click the
CEPH
to go into ceph mode.
2. Set the High-Resolution/Lateral Mode/exposure condition 80 kVp, 10 mA, and click the
Confirm
button.
3. Place the CEPH Phantom on the ear rod.
4. Attach 0.8 mm Copper Filter on X-Ray Tube.
5. Click the
READY
button and acquire and save the image according to the Console Software
guidance.
6. Run View16.exe and open the capture.raw file to measure the center position of the
Phantom.
Summary of Contents for A9 PHT-30CSS
Page 1: ...vatech A9 Technical manual Model PHT 30CSS Version 1 01 English...
Page 12: ...Page 11 of 244 1 2 1 Vertical Frame Part...
Page 22: ...Page 21 of 244 1 2 3 5 Generator DG 07E22T2...
Page 26: ...Page 25 of 244...
Page 27: ...Page 26 of 244...
Page 30: ...Page 29 of 244 1 3 3 Sensor Part Connection Diagram...
Page 31: ...Page 30 of 244 1 3 4 CEPH SENSOR Part Connection Diagram...
Page 32: ...Page 31 of 244 1 3 5 Connection Diagram Total assembly...
Page 43: ...Page 42 of 244 16 Loosen ten bolts as shown in the image below...
Page 112: ...Page 111 of 244 5 Click the V Vol button Click the Ready button...
Page 180: ...Page 179 of 244 10 Select 50 in the M box and click the center of the flat washer...
Page 209: ...Page 208 of 244 8 5 Appendix 8 5 1 Main MCU Board Parameters...
Page 244: ...vatech A9 www vatech com...