6. Click the
V Vol
button.
Y axis - reference range: 130 ~ 135
130 or less: Reassemble the bite block in the opposite direction of the
column (5 voxel value increase per cell)
135 or more: Reassemble the bite block in the direction of the column (5
voxel value decrease per cell)
X axis - reference range: 220 ~ 230
- 220 or less: move to the left based on the column
- Over 230: Move to the right based on the column
Angle setting
When moving to the right, Start angle: 0 ~ 10
When moving to the left, Start angle: 350 ~ 359
When the start angle is found, it is applied to the following folders.
C:\VCaptureSW\Acquisition\CBCT\WidePANO\REC\reconstruction.vxm
Change the value of the STARTANGLE parameter in the reconstruction.vxm
file
1mm per division
(5 voxel)
Summary of Contents for A9 PHT-30CSS
Page 1: ...vatech A9 Technical manual Model PHT 30CSS Version 1 01 English...
Page 12: ...Page 11 of 244 1 2 1 Vertical Frame Part...
Page 22: ...Page 21 of 244 1 2 3 5 Generator DG 07E22T2...
Page 26: ...Page 25 of 244...
Page 27: ...Page 26 of 244...
Page 30: ...Page 29 of 244 1 3 3 Sensor Part Connection Diagram...
Page 31: ...Page 30 of 244 1 3 4 CEPH SENSOR Part Connection Diagram...
Page 32: ...Page 31 of 244 1 3 5 Connection Diagram Total assembly...
Page 43: ...Page 42 of 244 16 Loosen ten bolts as shown in the image below...
Page 112: ...Page 111 of 244 5 Click the V Vol button Click the Ready button...
Page 180: ...Page 179 of 244 10 Select 50 in the M box and click the center of the flat washer...
Page 209: ...Page 208 of 244 8 5 Appendix 8 5 1 Main MCU Board Parameters...
Page 244: ...vatech A9 www vatech com...