Page
114
of
244
8. Enter the measured CT Number value and click the
Save
button.
9. When CT Number Calibration is completed, open
Reconstruction.vxm
and recover the
parameter as shown below.
• POSTPROCESSING = 0 → POSTPROCESSING =
-50
• UserCTNumCal = 0 → UserCTNumCal = 1
• Us
eArcCut= 0 → UseArcCut= 1
10. Capture the CT Number phantom and evaluate CT Number accuracy.
(refer to 5. CT Number Accuracy evaluation.)
※
CT Number standard value
Parameter
LSL(Lower Specification Limit)
USL(Upper Specification Limit)
Air
-1030
-900
Water
-20
20
FOV Info (CBCT)
Choose HD Tap and input CT Number value
measured at high resolution.