Page
180
of
244
11. Select
20
in the
M
box and make sure that the Ear-Rod guide pin is in the blue ROI area.
12. If the pin is out of the blue ROI area, do the steps that follow
4) Disassemble the top cover from CEPH Unit. Refer to
Chapter 2. Hardware Part
Service Guide
.
5) Loosen 2 wrench bolts and adjust the fixed block of EAR-ROD as shown below.
•
If turn the fixed block of EAR-ROD to the right on the picture below, the
Guide Pin of EAR-ROD will move from the right to the left on the picture
in the step11), and vice versa.
6) Exit
VAKCAP
. Acquire the PHANTOM image again and measure the Pixel Number at
the center of PHANTOM.
7) Repeat the procedures until the Ear-Rod is aligned correctly.
Summary of Contents for A9 PHT-30CSS
Page 1: ...vatech A9 Technical manual Model PHT 30CSS Version 1 01 English...
Page 12: ...Page 11 of 244 1 2 1 Vertical Frame Part...
Page 22: ...Page 21 of 244 1 2 3 5 Generator DG 07E22T2...
Page 26: ...Page 25 of 244...
Page 27: ...Page 26 of 244...
Page 30: ...Page 29 of 244 1 3 3 Sensor Part Connection Diagram...
Page 31: ...Page 30 of 244 1 3 4 CEPH SENSOR Part Connection Diagram...
Page 32: ...Page 31 of 244 1 3 5 Connection Diagram Total assembly...
Page 43: ...Page 42 of 244 16 Loosen ten bolts as shown in the image below...
Page 112: ...Page 111 of 244 5 Click the V Vol button Click the Ready button...
Page 180: ...Page 179 of 244 10 Select 50 in the M box and click the center of the flat washer...
Page 209: ...Page 208 of 244 8 5 Appendix 8 5 1 Main MCU Board Parameters...
Page 244: ...vatech A9 www vatech com...