11. Click each file in View 16 and check whether each median value meets that standard below:
Bright
Min
Max
Median value
Cal Point 1
4500
5500
Cal Point 2
9500
10500
[2-Point Bright Standard Value]
If the median value does not satisfy the standard, adjust
kVp0
and
kVp1
values from the
table below
(adjustment range:±10)
and repeat the bright calibration procedure.
Section
Values
kVp0
630
mA0
700
kVp1
750
mA1
910
The bright calibration date is saved at
C:\VCaptureSW\Acquisition\CBCT\WidePANO\CAL.
12. Create a new folder in the CAL folder and change the folder name to OPEN.
13. Copy these files to the C:\VCaptureSW\Acquisition\CBCT\WidePANO\CAL\OPEN.
Ensure to save all data in the OPEN folder after 2-point bright
calibration.
Summary of Contents for A9 PHT-30CSS
Page 1: ...vatech A9 Technical manual Model PHT 30CSS Version 1 01 English...
Page 12: ...Page 11 of 244 1 2 1 Vertical Frame Part...
Page 22: ...Page 21 of 244 1 2 3 5 Generator DG 07E22T2...
Page 26: ...Page 25 of 244...
Page 27: ...Page 26 of 244...
Page 30: ...Page 29 of 244 1 3 3 Sensor Part Connection Diagram...
Page 31: ...Page 30 of 244 1 3 4 CEPH SENSOR Part Connection Diagram...
Page 32: ...Page 31 of 244 1 3 5 Connection Diagram Total assembly...
Page 43: ...Page 42 of 244 16 Loosen ten bolts as shown in the image below...
Page 112: ...Page 111 of 244 5 Click the V Vol button Click the Ready button...
Page 180: ...Page 179 of 244 10 Select 50 in the M box and click the center of the flat washer...
Page 209: ...Page 208 of 244 8 5 Appendix 8 5 1 Main MCU Board Parameters...
Page 244: ...vatech A9 www vatech com...