4.3.5.
PANO Sensor: Bright Calibration
Stay outside of the X-ray room during Bright Calibration data achievement.
1. Run the
VACAP.exe
file (
Path
:
C:\VCaptureSW\Acquisition\Pano\WidePANO\VACAP.exe
)
2. On the dialog box, select Settings and follow the steps below to set exposure condition:
1) Go to
[BrightCalibration]
and enter
5
for NumPoints.
2) Check whether the tube voltage (kVp) and current (mA) meets the standards. If
necessary, enter voltage and current values according to the table below:
Acquisition conditions
5-point
exposure
condition
kVp0
620
mA0
600
kVp1
900
mA1
900
3) Click
Save
.
3. Send
[SPM_PSON]
to MCU to supply power to the PANO Sensor.
①
②
③