Page
69
of
244
9. Repeat
steps 7
to
8
until you acquired four bright calibration data files.
10. Click the
V
to open
View16
.
11. Click each file in View 16 and check whether each median value satisfies the standards in
the table below:
Bright
Min
Max
Median
value
Cal Point 1
80
120
Cal Point 2
150
700
Cal Point 3
800
1800
Cal Point 4
2000
2500
[4-Point Bright Calibration Standard Value]
If the median value falls outside of the standard, adjust the kVp0 and kVp1 values and
repeat the calibration procedure. (Adjustment range:
±
10)
12. Copy the saved data from the OPEN folder and paste them to the CAL folder. After this step,
the CAL folder must contain 6 files.
Point 1
Point 2
Point 3
Point 4