6. After confirming of position, capture the image to measure DAP.
7. Run the Console Software, and set as below and capture the image.
8. Measure mGy and check if it meets the standard as follows. If not, adjust mA and re-capture.
※
Background level standard value
Mode
exposure condition
mGy
Background Level
Man (High resolution)
95kVp
8.5 (±0.1)
7000 ~ 11500
Man (Green)
80kVp
5.4 (±0.1)
2000 ~ 5500
3. Choose the High resolution.
2. Choose the FOV: 8x8
4. Choose the Standard.
5. Choose the
Man.
1. Choose the CBCT.
6. Choose the
Normal.
7. Click the Confirm button.