Page
67
of
244
2.3.5.2
How to Perform 4-Point Bright Calibration:
To perform 4-point bright calibration:
1. Run the VAKAP.exe file. (
Path
:
C:\VCaptureSW\Acquisition\CBCT\WidePANO\VACAP.exe
)
2. On the dialog box, select
Settings
and follow the steps below to set exposure condition:
1) Go to
[BrightCalibration]
and enter
4
for the NumPoints
2) Go to the 4-point section and check whether the tube voltage(kVp) and current(mA) meets
the standards. If necessary, enter each value according to the table bleow:
Section
Values
kVp0
570
mA0
700
kVp1
880
mA1
950
3) Go to the 2-point section and insert “//” before all data
(example: // kVp0=630)
4) Click
Save
.
①
②
③
④