3.6.3.
Test Method
1. Put the Phantom jig on the unit and make the phantom jig level by using bubble level and three
align pins.
2. Put the S&C Phantom on the phantom jig.
3. Run the Console Software and click the CBCT button to go into CBCT mode.
4. Choose the FOV 8x8/Adult/Normal/High Resolution/Standard as the image option and click the
Confirm button.
5. Press the Ready button and acquire an image.
6. When the reconstruction is completed, save the CT folder name in the
C:\VCaptureSW\ImageOutput folder as CT_S&C Phantom.
7. Open the saved image through 3D Viewer.
8. Go to the Axial view in full-screen mode and adjust the brightness and contrast of the image.
9. Make sure that you can see the minimum size of Air hole and PE(PolyEthylene) cylinder at the
distance of 50 cm(20 inches) from the monitor.
3.6.4.
Test Result
Parameter
LSL
Measurement value
USL
Pass/Fail
High Contrast Resolution
-
1.0 mm
Summary of Contents for A9 PHT-30CSS
Page 1: ...vatech A9 Technical manual Model PHT 30CSS Version 1 01 English...
Page 12: ...Page 11 of 244 1 2 1 Vertical Frame Part...
Page 22: ...Page 21 of 244 1 2 3 5 Generator DG 07E22T2...
Page 26: ...Page 25 of 244...
Page 27: ...Page 26 of 244...
Page 30: ...Page 29 of 244 1 3 3 Sensor Part Connection Diagram...
Page 31: ...Page 30 of 244 1 3 4 CEPH SENSOR Part Connection Diagram...
Page 32: ...Page 31 of 244 1 3 5 Connection Diagram Total assembly...
Page 43: ...Page 42 of 244 16 Loosen ten bolts as shown in the image below...
Page 112: ...Page 111 of 244 5 Click the V Vol button Click the Ready button...
Page 180: ...Page 179 of 244 10 Select 50 in the M box and click the center of the flat washer...
Page 209: ...Page 208 of 244 8 5 Appendix 8 5 1 Main MCU Board Parameters...
Page 244: ...vatech A9 www vatech com...