82
minute, it returns 1; when the unit is time, it returns 2. The command is invalid
when the output is turned on.
6.3.5.25
:FUNC:SD
Command Syntax: :FUNCtion:SurgeDrop:MANUal ON or :FUNCtion:SurgeDrop:MANUal 1
Shorthand for command: :FUNC:SD:MANU ON or :FUNC:SD:MANU 1
Command description: This command is used to enable the surge drop function in manual mode. The
command is invalid when the output is turned on.
Command Syntax: :FUNCtion:SurgeDrop:MANUal OFF or :FUNCtion:SurgeDrop:MANUal 0
Shorthand for command: :FUNC:SD:MANU OFF or :FUNC:SD:MANU 0
Command description: This command is used to turn off the surge drop function in manual mode. The
command is invalid when the output is turned on.
Command Syntax: :FUNCtion:SurgeDrop:MANUal?
Shorthand for command: :FUNC:SD:MANU?
Command description: This command is used to query the state of the surge drop in manual mode.
Command Syntax: :FUNCtion:SurgeDrop:PROGram ON or:FUNCtion:SurgeDrop:PROGram 1
Shorthand for command: :FUNC:SD:PROG ON or:FUNC:SD:PROG 1
Command description: This command is used to enable the surge drop at the current step in the
programmable mode. The command is invalid when the output is turned on.
Command Syntax: :FUNCtion:SurgeDrop:PROGram OFF or:FUNCtion:SurgeDrop:PROGram 0
Shorthand for command: :FUNC:SD:PROG OFF or:FUNC:SD:PROG 0
Command description: This command is used to turn off the surge drop at the current step in the
programmable mode. The command is invalid when the output is turned on.
Command Syntax: :FUNCtion:SurgeDrop:PROGram?
Shorthand for command: :FUNC:SD:PROG?
Command description: This command is used to query the state of the surge drop at the current step in
the programmable mode. The command is invalid when the output is turned on.
Command Syntax: :FUNCtion:SurgeDrop:VOLT:MANUal <value>
Shorthand for command: :FUNC:SD: VOLT:MANU <value>
Command description: This command is used to set the voltage value of the surge drop in manual mode.
<value> indicates the value of the glitch trap voltage, ranging from 0.0 to 300.0V.
Command Syntax: :FUNCtion:SurgeDrop:VOLT:MANUal?
Shorthand for command: :FUNC:SD: VOLT:MANU?
Command description: This command is used to query the voltage value of the surge drop in manual
mode. Returns <value> with a value equal to 0.0~300.0V.
Command Syntax: :FUNCtion:SurgeDrop:VOLT:PROGram <value>