![Tonghui Electronics TH7105 Operation Manual Download Page 126](http://html1.mh-extra.com/html/tonghui-electronics/th7105/th7105_operation-manual_1126456126.webp)
121
1~32
0x10
0x00
0x2B
0x00
0x01
0x02
0x00
0x00 or
0x01 or
0x02
Calculated based
on the first 8 bytes
3)
Return Instruction Format
6.4.2.43
Address 44 (Delay Time)
1)
Command Description
This command is used to set the delay determination time of the current step in the programmed mode.
When the unit is second, the range is 0.1~999.9s; when the unit is minute, the range is 0.1~999.9m; when
the unit is time, the range is 0.1~999.9h. Command is invalid when output is turned on
.
2)
Send Instruction Format
Instrument
address
Function
code
High
address
Low
address
High
bytes
Low
bytes
Number
of data
Data byte 1~4
CRC
Low
CRC
High
1~32
0x10
0x00
0x2C
0x00
0x02
0x04
0.1~999.9, 4 bytes
stored in float
point type
Calculated based on
the first 11 bytes
3)
Return Instruction Format
6.4.2.44
Address 45 (Test Time)
1)
Command Description
This command is used to set the test time of the current step in the programmed mode. When the unit is
second, the range is 0.1~999.9s; when the unit is minute, the range is 0.1~999.9m; when the unit is time,
the range is 0.1~999.9h. Command is invalid when output is turned on
.
2)
Send Instruction Format
Instrument
address
Function
code
High
address
Low
address
High
bytes
Low
bytes
Number
of data
Data byte 1~4
CRC
Low
CRC
High
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x10
0x00
0x2B
0x00
0x01
Calculated based on the first 6
bytes
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x10
0x00
0x2C
0x00
0x04
Calculated based on the first 6
bytes