6.4.3.45
Address 45 (Test Time)
1)
Command Description
This command is used to query the test time of the current step in the programmed mode.
2)
Send Instruction Format
3)
Return Instruction Format
The return instruction data range is from 0.1 to 999.9. When the time unit is second, the value is
0.1~999.9s; when the time unit is minute, the value is 0.1~999.9m; when the time unit is time, the value
is 0.1~999.9h.
6.4.3.46
Address 46 (Rise Time)
1)
Command Description
This command is used to query the rise time of the current step in the programmed mode.
2)
Send Instruction Format
3)
Return Instruction Format
Return instruction data range is 0.1~999.9s.
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x03
0x00
0x2D
0x00
0x02
Calculated based on the first 6
bytes
Instrument
address
Function Code
Number of Data
Data Byte 1~4
CRC Low
CRC High
1~32
0x03
0x04
0.1~999.9, 4 bytes
stored in float point
type
Calculated based on the first 7 bytes
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x03
0x00
0x2E
0x00
0x02
Calculated based on the first 6
bytes
Instrument
address
Function Code
Number of Data
Data Byte 1~4
CRC Low
CRC High
1~32
0x03
0x04
0.1~999.9, 4 bytes
stored in float point
type
Calculated based on the first 7 bytes