144
3)
Return Instruction Format
Return instruction data byte range is 0~59.
6.4.3.26
Address 26 (Timer Hour)
1)
Command Description
This command is used to query the number of hours of timer in the manual mode.
2)
Send Instruction Format
3)
Return Instruction Format
Return instruction data byte range is 0~99.
6.4.3.27
Address 27 (Memory Sequence)
1)
Command Description
This command is used to query the serial number of the memory sequence in the programmed mode.
2)
Send Instruction Format
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x03
0x00
0x19
0x00
0x01
Calculated based on the first 6
bytes
Instrument
address
Function
Code
Number of
Data
Data Byte1
Data Byte2
CRC Low
CRC High
1~32
0x03
0x02
0x05
0~59
Calculated based on the first 5
bytes
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x03
0x00
0x1A
0x00
0x01
Calculated based on the first 6
bytes
Instrument
address
Function
Code
Number of
Data
Data Byte1
Data Byte2
CRC Low
CRC High
1~32
0x03
0x02
0x00
0~99
Calculated based on the first 5
bytes
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High