Return Instruction Format
6.4.2.57
Address 58 (Test Result)
1)
Command Description
This command is used to set the result display mode in the programmed mode. Command is invalid when
output is turned on
.
The instruction data byte bit corresponds to the function of the instrument as shown
below
:
Data Byte Result Display Mode
0x00
This command is used to set the result display mode to LAST in
programmed mode.
0x01
This command is used to set the result display mode to ALL in
programmed mode.
0x02
This command is used to set the result display mode to P/F in
programmed mode.
2)
Send Instruction Format
Instrument
address
Function
code
High
address
Low
address
High
bytes
Low
bytes
Number
of data
Data
byte 1
Data byte
2
CRC
Low
CRC
High
1~32
0x10
0x00
0x3A
0x00
0x01
0x02
0x00
0x00 or
0x01 or
0x02
Calculated based
on the first 8 bytes
3)
Return Instruction Format
6.4.2.58
Address 59 (Surge Drop)
1)
Command Description
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x10
0x00
0x39
0x00
0x01
Calculated based on the first 6
bytes
Instrument
address
Function
Code
High Address
Low Address
High bytes
Low bytes
CRC Low
CRC High
1~32
0x10
0x00
0x3A
0x00
0x01
Calculated based on the first 6
bytes